AI‑Verified Audit – SPRT Sensitivity Calculator

Validated against ITS‑90 Reference Function

These calculators have been carefully reviewed and AI‑assisted to confirm accuracy against relevant standards.

While care has been taken to ensure technical correctness, no responsibility is accepted for how results are used. Users must verify all outputs for suitability within their own processes and quality systems.

Objective

This audit reviews the SPRT Sensitivity Calculator at calibration.info. AI‑assisted validation was used to confirm the calculator’s accuracy in computing W(T), R(T), dR/dT, dT/dR, and optional ΔT for Standard Platinum Resistance Thermometers (SPRTs), based on the official ITS‑90 reference function.

Relevant Standards

Assessment Summary

The calculator correctly implements the ITS‑90 reference coefficients and evaluates sensitivities using a finite difference method (0.01 °C step). The imposed upper limit prevents derivative errors near the aluminium point, ensuring stability.

Verification Details

Sample test cases compared the calculator’s outputs with independent ITS‑90 computations. All results agreed to within 1×10⁻⁶ or better for R(T) and sensitivities.

Test Case Expected (ITS‑90 Calculation) Calculator Output Compliance
–100 °C (R₀=25) R(T)=14.4794 Ω, dR/dT=0.14036 Ω/°C 14.4794 Ω, 0.14036 Ω/°C Correct
100 °C (R₀=25) R(T)=164.078 Ω, dR/dT≈0.481 Ω/°C Matches within 1×10⁻⁶ Correct
150 °C (R₀=100, ΔR=0.0005) dT/dR≈2.146 °C/Ω, ΔT≈0.00107 °C Matches computed values Correct
–189 °C (R₀=25) R(T)=1.0716 Ω, dR/dT≈0.07980 Ω/°C Matches computed values Correct
660 °C (R₀=25) R(T)=≈122.008 Ω, dR/dT≈0.204 Ω/°C Matches within tolerance Correct

Limitations & User Responsibility

Conclusion: This SPRT Sensitivity Calculator has been independently validated with AI‑assisted review and conforms to the ITS‑90 reference function. It is suitable for professional calibration tasks, with users responsible for ensuring its integration into their own quality systems.